15.3 ABG, Monitoring & QC Reliability
NBRC Registered Pulmonary Function Technologist RPFT
Track your progress as you read
Finished sections get a check mark.
1 Introduction & Exam Overview
1.1 Current Exam Facts 1.2 Content Outline Domains & Study Plan
2 Spirometry Instrumentation
2.1 Spirometer Types, Setup & Maintenance 2.2 Spirometer Calibration & Linearity 2.3 Spirometer Troubleshooting
3 Blood Gas & Gas Analyzers
3.1 Blood Gas Analyzer Setup, Maintain, Calibrate 3.2 Gas Analyzers & Water/Gas Absorbers 3.3 ABG & Gas Analyzer Troubleshooting and QC
4 Body Box, Exercise & Monitoring Equipment
4.1 Body Plethysmograph Setup & Calibration 4.2 Exercise, Field Walking & ECG Equipment 4.3 Pressure Devices, Valves & Gas Delivery Systems
5 Ancillary Equipment & Infection Control
5.1 Body Habitus, Sampling & Aerosol Equipment 5.2 Monitors & Emergency Management Equipment 5.3 Infection Control Materials & Methods
6 Quality Control Systems
6.1 Mechanical QC Devices 6.2 Biological Controls & Proficiency Testing 6.3 Laboratory Quality Management & SOPs
7 Patient Preparation & Safety
7.1 Body Habitus Measurements & Estimates 7.2 Patient Safety, Precautions & Adverse Events
8 Spirometry Procedures
8.1 Selecting Spirometry Protocols & Equipment 8.2 Performing Standard & Position-Variant Spirometry 8.3 Evaluating Spirometry Acceptability & Repeatability 8.4 Bronchodilator Response Studies
9 Static Lung Volumes
9.1 Gas Dilution Lung Volume Methods 9.2 Body Plethysmography Lung Volumes 9.3 Evaluating Lung Volume Validity
10 Diffusing Capacity (DLCO)
10.1 DLCO Protocol & Equipment Selection 10.2 Performing the DLCO Maneuver 10.3 Evaluating DLCO Validity
11 Blood Gas Sampling & Analysis Procedures
11.1 Arterial & Capillary Blood Sample Collection 11.2 Performing Blood Gas Analysis 11.3 CO-oximetry / Hemoximetry Procedures
12 Airway Responsiveness, Resistance & Muscle Strength
12.1 Bronchial Provocation Studies 12.2 Airways Resistance & Conductance by Plethysmography 12.3 Respiratory Muscle Strength (MIP/MEP/CPF)
13 Exercise, Oxygen, Home Testing & Education
13.1 Field Walking Tests (6MWT, Shuttle) 13.2 Monitored Exercise Testing 13.3 Oxygen Assessment & Titration 13.4 Home Testing Instruction & Patient Education
14 Calculations & Reference Ranges
14.1 Spirometry Calculations & Reference Equations 14.2 Lung Volume & DLCO Calculations/Corrections 14.3 Blood Gas & CO-oximetry Derived Calculations 14.4 Exercise, Oximetry & Special Test Calculations
15 Evaluating Result Reliability
15.1 Spirometry & Lung Volume Reliability 15.2 DLCO & Exercise Test Reliability 15.3 ABG, Monitoring & QC Reliability
16 Clinical Implications & Reporting
16.1 Pattern Recognition: Obstruction, Restriction, Mixed, Severity 16.2 Clinical Implications: DLCO, ABG, Airway Response, Exercise & Lab Context
15.3 ABG, Monitoring & QC Reliability
Key Takeaways Explain This Topic AI
Pre-analytical ABG errors (air bubbles, delay/metabolism, excess heparin) bias PO2/PCO2/pH before the analyzer ever runs—reliability starts at the syringe.
Air contamination drives PO2 toward room-air PO2 (~150 mmHg at sea level on air) and lowers PCO2; never treat bubble-tainted gases as patient truth.
Pulse oximetry is unreliable with significant COHb or MetHb, dark nail coatings, and poor perfusion—use co-oximetry/ABG pathways when stems demand accuracy.
Exercise BP and ECG artifacts (motion, loose leads, wrong cuff) must be recognized so false “hypertension” or arrhythmia does not drive unsafe or wrong conclusions.
Out-of-range QC means do not release patient results; safety events and lab QC status are Domain III.B reliability inputs (including III.B.15–16).
Last updated: August 2026
Reliability Beyond the PFT Booth
Domain III.B is not limited to spirometry graphs. Blood gases, noninvasive monitors, exercise hemodynamics, and laboratory QC/safety data all feed the same question: Are these results reliable enough to act on? Tasks spanning III.B.1, III.B.2, III.B.9–11, and III.B.15–16 reward technologists who catch pre-analytical , monitoring , and QC-system failures before a report becomes a clinical decision.
Pre-Analytical ABG Errors (High-Yield Mechanics)
Most “mystery” blood-gas numbers are ruined before analysis .
Air bubbles
Room air has high PO2 (~150 mmHg at sea level, dry air approximation) and near-zero PCO2 relative to arterial blood.
- If sample has air bubbles… Typical directional bias
- PO2 Moves toward room-air PO2 (raises low PO2; can lower very high PO2 on high FiO2)
- PCO2 Falls toward air
- pH Often rises as PCO2 falls (respiratory alkalosis artifact)
Reliability action: Expel bubbles immediately; analyze promptly; reject visibly aerated samples when policy requires. Do not “interpret around” a bubble.
Delay and metabolism
In a closed syringe, cells continue metabolism if analysis is delayed and temperature is not controlled.
- Delay effect (typical) Direction
- PO2 Tends to fall (O2 consumption)
- PCO2 Tends to rise
- pH Tends to fall
Icing/rapid analysis rules exist to preserve reliability . A stem with a sample left warm for a long time before running is a pre-analytical reliability failure—not a new metabolic acidosis diagnosis until redrawn correctly.
Excess heparin / dilution
Liquid heparin can dilute the sample and alter PCO2/HCO3-related values; improper anticoagulant technique also risks clots that damage analyzers and delay processing.
- Error Reliability impact
- Excess liquid heparin Dilutional bias; especially problematic for PCO2/HCO3 narratives
- Clotted sample May be unanalyzable; delays → secondary metabolic change
- Wrong syringe type / uncapped sample Gas exchange with air; evaporation
- Other pre-analytical reliability threats
- Venous or mixed sample labeled as arterial → falsely low PO2, higher PCO2 pattern
- Patient state mislabeled (FiO2, temperature, site) → correct numbers, wrong clinical frame
- Plastic syringe storage too long → gas diffusion through plastic over time
Exam habit: When ABG values contradict clinical SpO2/ABG history abruptly, audit the syringe path before rewriting the patient’s physiology.
Pulse Oximetry Reliability Limits
SpO2 estimates oxyhemoglobin fraction using limited wavelengths. It is not a co-oximeter.
Condition SpO2 reliability problem Better path
Elevated COHb SpO2 tends to read high relative to true oxygen-carrying status ; cannot distinguish COHb Co-oximetry (COHb, O2Hb)
- MetHb SpO2 often trends toward mid-range (~85%) artifactually Co-oximetry (MetHb)
- Dark nail polish / opaque coverings Optical interference, low signal Remove polish or alternate site
- Poor perfusion / cold / shock Unstable or false readings Warm site, alternate probe, ABG if needed
- Motion Spurious desaturation or spikes Confirm pulse rate agreement
- Ambient light / probe malposition Noise Shield/reseat probe
Reliability rule: SpO2 is excellent for continuous trending when signal quality is good and dyshemoglobins are not in play. It is unreliable as the sole truth for CO exposure, methemoglobinemia, or low-perfusion states highlighted in exam vignettes.
BP and ECG Artifact During Exercise (III.B monitoring reliability)
Exercise introduces motion that invents pathology on paper.
Blood pressure artifact
- Artifact source False appearance Mitigation
- Talking / isometric grip on rails Spurious BP rise Standardize arm position; quiet measurement windows
- Wrong cuff size / loose cuff High or erratic BP Correct cuff; secure
- Motion during inflation Failed or nonsense readings Manual confirmation; timing at stage end
- Arm below/above heart level Hydrostatic error Heart-level transducer/cuff
Do not escalate care or stop tests solely on a single motion-contaminated BP without a repeat under better conditions—unless symptoms/ECG demand safety stop.
ECG artifact
- Artifact Looks like Check
- Loose electrodes / sweat Baseline wander, fake “ischemia” Re-prep skin; secure leads
- Motion / cable tug Spikes mimicking PVCs Correlate pulse; inspect raw
- 60-Hz interference Thick baseline Power/cable routing
- Wrong lead placement Axis/morphology errors Verify landmarks
Reliability means distinguishing noise from arrhythmia . True ectopy that is pulse-confirmed differs from cable slap that vanishes when the arm rests.
QC Data Reliability: The Hard Stop
Instrument QC is a release gate , not paperwork decoration.
- QC status Patient-result action
- Controls in range (and other system checks OK) May analyze/release per SOP
- Control out of range / failed multi-rule Do not release patient results from that analytic system
- After corrective action Re-QC; only then resume patients
- Marginal / trending toward failure Investigate early; protect reliability
Classic RPFT item: Analyzer QC fails; three ABGs already drawn. Correct answer path is hold results , troubleshoot/recalibrate/re-control per SOP, repeat analysis only when QC passes—not “report with a disclaimer footnote” while controls are red.
The same logic applies to spirometers, gas analyzers, and co-oximeters: out-of-control measurement systems produce unreliable patient data regardless of how cooperative the patient was.
Safety Data and Lab QC as Reliability Inputs (III.B.15–16)
Reliability is also organizational:
Adverse events / near misses during testing (syncope, severe desaturation, needle sticks) affect whether a dataset was obtained under valid physiologic and safety conditions and whether the process is safe to continue.
Competency, SOP version, proficiency testing, and equipment maintenance logs tell you whether the lab’s measurement system deserves trust today.
Critical value notification and amended-report pathways protect patients when a previously released result is later found unreliable.
If safety equipment failed mid-exercise or an emergency interrupted stages, remaining numeric fragments may be technically incomplete —document and limit interpretation rather than forcing a full CPET narrative.
Scenario Table for Rapid Exam Recognition
- Stem clue Likely reliability issue Best next step
- ABG PO2 “normalizes” toward 150 with low PCO2 after difficult draw Air bubbles Reject/redraw; expel air
- ABG pH falls, PCO2 rises after long warm delay Metabolism Rapid analysis/icing per SOP; redraw if needed
- SpO2 100% but patient smoke-exposed and COHb high SpO2 blind to COHb Co-oximetry
- SpO2 ~85% with chocolate-brown blood suspicion MetHb Co-oximetry
- Exercise BP 220/40 with motion Artifact Repeat with technique fix
- Blood gas QC level 2 fails System unreliability No patient release until QC OK
- Walk SpO2 crash with HR mismatch Oximeter artifact Reseat; corroborate
- Integrated Judgment Workflow
ABG: syringe integrity → air → delay → heparin/clot → label FiO2/site → then analyzer QC status.
SpO2: signal quality → pulse agreement → nail/perfusion/dyshemoglobin risks.
Exercise monitors: BP cuff mechanics + ECG lead integrity before clinical labels.
QC/safety: red control or unsafe/incomplete acquisition → stop release , fix system, retest as appropriate.
Only clean, controlled, well-documented data proceed to III.C clinical implications .
Domain III.B closes the loop between instruments (Domain I), procedures (Domain II), and trustworthy information. An RPFT’s professional product is not a stack of numbers—it is a reliable dataset that clinicians can defend.